Catalog Number 03666794001
Device Problem Circuit Failure (1089)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/29/2020
Event Type  malfunction  
Manufacturer Narrative
The customer's device was requested for investigation.
 
Event Description
The initial reporter complained of a display issue with a cc-xs meter, which could cause misinterpretation of results.The customer stated the meter¿s display is very faint and hard to see, and there might be missing segments.The customer mentioned he has placed new batteries within the meter but the issue persisted.The customer performed a display check, and all the segments were present but faint.Also, the values of "88.8 looked more like 66.6." after the display check, the meter flashed "12/31/00" and "set," which the customer stated was very faint.No misinterpretation of results were reported.
 
Manufacturer Narrative
The customer's meter was received for investigation.The meter¿s circuit board was tested for damage or contamination and it was found the printed circuit board (pcb) was contaminated.The root cause is determined to be contamination of the contacts due to improper handling or maintenance by the customer.Medwatch fields d9 and h3 were updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key10434222
MDR Text Key203832780
Report Number1823260-2020-02050
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 09/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number03666794001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2020
Initial Date Manufacturer Received 07/29/2020
Initial Date FDA Received08/20/2020
Supplement Dates Manufacturer Received07/29/2020
Supplement Dates FDA Received09/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-