• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STANMORE IMPLANTS WORLDWIDE METS DISTAL FEMUR - FEMORAL STEM; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STANMORE IMPLANTS WORLDWIDE METS DISTAL FEMUR - FEMORAL STEM; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number UNK_STM
Device Problems Unstable (1667); Loss of Osseointegration (2408)
Patient Problems Pain (1994); Injury (2348); Inadequate Osseointegration (2646); Cancer (3262)
Event Date 07/24/2020
Event Type  Injury  
Manufacturer Narrative
An investigation is being performed in an attempt to identify the cause of the event.Should additional information become available it will be reported in a supplemental report.Device not returned.
 
Event Description
Patient has presented with an unstable distal femoral replacement.Loosened stanmore mets distal femur.Patient specific prescription form notes: "would prefer an adapter that would allow the stanmore dfr to link up with a [competitor] spindle.If that is impossible, we will require a press-fit stem and also potentially a total femur as a backup.".
 
Event Description
Patient has presented with an unstable distal femoral replacement.Loosened stanmore mets distal femur.Patient specific prescription form notes: "would prefer an adapter that would allow the stanmore dfr to link up with a [competitor] spindle.If that is impossible, we will require a press-fit stem and also potentially a total femur as a backup.".
 
Manufacturer Narrative
Reported event: an event regarding loosening involving a mets distal femur replacement, femoral stem was reported.The event was confirmed by x ray review.Method & results: device evaluation and results: not performed as product was not returned.Clinician review: a review of the provided x-rays by clinical consultant indicated: the implant in situ was for a mets distal femoral replacement which was inserted in 2017.The surgeon reported that the implant has loosened.The ct scan provided shows that the bone had massive bone resorption and remodelling, and there was severe radiolucency around the stem.Therefore, the radiographic review can confirm the reason for revision.Device history review: could not be performed as lot / batch code information was not provided.Complaint history review: could not be performed as lot/ batch code information was not provided.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, pathology reports, progress notes, and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METS DISTAL FEMUR - FEMORAL STEM
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
STANMORE IMPLANTS WORLDWIDE
210 centennial avenue
centennial park, elstree
borehamwood WD6 3 SJ
UK  WD6 3SJ
MDR Report Key10434944
MDR Text Key203832616
Report Number3004105610-2020-00131
Device Sequence Number1
Product Code KRO
Combination Product (y/n)N
PMA/PMN Number
K121029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 11/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNK_STM
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/24/2020
Initial Date FDA Received08/20/2020
Supplement Dates Manufacturer Received11/06/2020
Supplement Dates FDA Received11/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age65 YR
Patient Weight77
-
-