Model Number N/A |
Device Problem
Power Problem (3010)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/06/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Evaluation and investigation is in process.Once the investigation is complete, a supplemental report will be filed.
|
|
Event Description
|
It was reported that during surgery the electric dermatome would not stay on once powered on.No additional graft was required.The delay was less than 15 minutes during surgery.No adverse events were reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This follow up report is being submitted to report additional information.Review of the most recent repair record determined the motor was malfunctioning.The motor was replaced and resolved the reported issue.Review of the previous repair record identified unrelated repairs to the reported event.The device was tested and found to be functioning to specifications prior to release to the customer.Device is used for treatment.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
|
|
Event Description
|
No additional information.
|
|