MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET LEAD DOUBLE BEND, 86 CM; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET LEAD DOUBLE BEND, 86 CM; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1457Q/86
Device Problems Device Dislodged or Dislocated (2923); No Pacing (3268)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/10/2020
Event Type  Injury  
Manufacturer Narrative
This product is registered as a combination product.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient presented to the hospital experiencing blood loss due a gastrointestinal bleed causing a patient fall.Upon interrogation, it was noted that the patient experienced loss of pacing.An x-ray was performed and the left ventricular - lv lead was found to be dislodged.It was noted that lv lead dislodged due to the fall the patient experienced.The lead was explanted and replaced.The patient was in stable condition afterwards.
 
Manufacturer Narrative
Analysis was normal.No anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET LEAD DOUBLE BEND, 86 CM
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key10438159
MDR Text Key203888806
Report Number2017865-2020-12013
Device Sequence Number1
Product Code OJX
UDI-Device Identifier05414734510189
UDI-Public05414734510189
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2023
Device Model Number1457Q/86
Device Catalogue Number1457Q/86
Device Lot NumberA000093573
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/10/2020
Initial Date FDA Received08/21/2020
Supplement Dates Manufacturer Received09/21/2020
Supplement Dates FDA Received09/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-