Model Number UL400
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Needle Stick/Puncture (2462)
Event Date 07/24/2020
Event Type  Injury  
Event Description
On (b)(6) 2020, neotract was informed of a physician who experienced difficulty in retracting the needle during a prostatic urethral lift (pul) procedure.It was reported that the device was successfully removed without implant placement; however, the needle remained extended in a deployed state.The procedure was successfully completed without further issues.There was no patient harm or injury reported.It was reported that while preparing the device to return for investigation, the needle pierced the bag and scratched the hand of a neotract representative present for the case.The appropriate incident report and blood tests were completed per hospital protocol for both the patient and the employee.All tests were favorable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEOTRACT UROLIFT SYSTEM UL400
Type of Device
UROLIFT SYSTEM
Manufacturer (Section D)
NEOTRACT, INC.
4155 hopyard road
pleasanton CA 94588
Manufacturer (Section G)
NEOTRACT, INC.
4155 hopyard road
pleasanton CA 94588
Manufacturer Contact
brian gall
4155 hopyard road
pleasanton, CA 94588
9253296547
MDR Report Key10438551
MDR Text Key211010350
Report Number3005791775-2020-00035
Device Sequence Number1
Product Code PEW
UDI-Device Identifier00814932020001
UDI-Public00814932020001
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K173087
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberUL400
Device Catalogue NumberAN00156
Device Lot NumberP42512
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/24/2020
Initial Date FDA Received08/21/2020
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-