Model Number FT4 G3
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/13/2020
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).The customer's calibration and qc data were requested but not provided.The observed differences in ft4 values generated with the roche assay and the accursed assay are most likely caused by differences in the overall setups of the assays, the antibodies used, differences in reference materials, and the differences in the standardization methodology used.The investigation did not identify a product problem.The cause of the event could not be determined.(b)(6).
 
Event Description
The initial reporter received questionable elecsys tsh assay and elecsys ft4 iii assay results for one patient tested on a cobas 8000 e 801 module.The e 801 module serial number was (b)(4).The customer reported out the results to a physician who asked for re-measurement of the sample.The customer performed repeat testing on an accursed analyzer, and the customer provided the patient¿s sample for investigation where it was tested with an abbott architect.This medwatch is for ft4.Refer to the medwatch with patient identifier (b)(6) for the tsh assay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III ASSAY
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key10439742
MDR Text Key204046926
Report Number1823260-2020-02070
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFT4 G3
Device Catalogue Number07976887190
Device Lot Number460793
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/31/2020
Initial Date FDA Received08/21/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age59 YR
-
-