• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD UNKN BIRMINGHAM HIP RESURFACING (BHR) CUP (UNKN TYPE); PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD UNKN BIRMINGHAM HIP RESURFACING (BHR) CUP (UNKN TYPE); PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number UNKNOWN
Device Problems Compatibility Problem (2960); Patient Device Interaction Problem (4001)
Patient Problems Cyst(s) (1800); Foreign Body Reaction (1868); Failure of Implant (1924); Pain (1994); Metal Related Pathology (4530)
Event Date 08/20/2019
Event Type  Injury  
Event Description
It was reported that, after a bhr construct had been implanted in a left hip on (b)(6) 2008, the patient experienced, as consequence of a failed left total hip replacement, pain and poor function; cobalt toxicity; pseudotumor with involvement of deep capsular structures and external rotators; atrophy and necrosis of tissue and fluid in hip joint.A revision surgery was performed on (b)(6) 2019 to treat the adverse event.The patient outcome is unknown.
 
Manufacturer Narrative
It was reported that left hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.As no device part and batch numbers were provided for investigation, a complaint history review, manufacturing record review and device labelling / ifu review could not be performed.If more information is received, this investigation will be reopened.A risk management review was performed.No additional risks were identified as result of the reported event and no further actions are required at this time.Smith and nephew have not received the device/adequate materials to fully evaluate the complaint, but if additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Manufacturer Narrative
H3, h6: it was reported that left hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.As no device part and batch numbers were provided for investigation, a complaint history review, manufacturing record review and device labelling/instructions for use review could not be performed.If more information is received, this investigation will be reopened.A risk management review was performed.No additional risks were identified as result of the reported event and no further actions are required at this time.The available medical documents were reviewed.The revision operative report does not note findings consistent with the complaint description of pseudotumor with involvement of deep capsular structures and external rotators; atrophy and necrosis of tissue and fluid in hip joint.Although the reported pain, femoral loosening, osteolysis and metal staining may be consistent with metal debris; without the supporting lab/pathology results, imaging, and/or the analysis of the explanted components, the root cause of the reported clinical reactions cannot be confirmed and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN BIRMINGHAM HIP RESURFACING (BHR) CUP (UNKN TYPE)
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa warwickshire
UK 
MDR Report Key10440339
MDR Text Key203987748
Report Number3005975929-2020-00318
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup,Followup
Report Date 01/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/03/2020
Initial Date FDA Received08/21/2020
Supplement Dates Manufacturer Received09/14/2020
01/28/2021
Supplement Dates FDA Received09/17/2020
01/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-