A field service engineer (fse) was at the customer site to address the reported issue.The fse confirmed the error on the error log; however, could not reproduce the issue.The fse noticed dried serum buildup on the sample nozzle and replaced the sample nozzle.The fse then primed the system and verified nozzle alignments.Impasse sense was then run and passed.Daily check and controls performed also passed.The aia-360 instrument returned to normal operation.No further field action required.The suspect nozzle assembly was returned to tosoh instrument service center for investigation.A visual inspection revealed no shipping damage.Functional testing was performed by installing the nozzle assembly in an aia-360r test bed instrument.The nozzle was aligned per the compound action.Controls were run to check for clogged nozzle.A 10-point precision test was performed on level 3 control without any error.The reported issue could not be replicated.The nozzle assembly functioned as expected.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from 29jun2019 to aware date 29jul2020.One other similar complaint was identified during the search period.The aia-360 operator's manual states the following: [4017] spec.Sy clog detected.Description: specimen clog was detected.Troubleshooting: the item was not assayed.Repeat assay.The probable cause of the issue could not be determined.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|