A field service engineer (fse) was at the customer site to address the reported issue.The fse confirmed and replicated the issue by running qc on the instrument.The fse replaced the sample nozzle; however, the error persisted.The pressure sensor was then replaced and the impasse test result was acceptable but on the high end.Next, the solenoid valve on the sample syringe was replaced which brought the impasse test result near the middle, and the issue was resolved.Instrument validation was performed via quality control (qc); results were within the published ranges.The aia-360 instrument returned to normal operation.No further field action required.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from (b)(6) 2019 to aware date (b)(6) 2020.No other similar complaints were identified during the search period.The aia-360 operator's manual states the following: [4017] spec.Sy clog detected.Description: specimen clog was detected.Troubleshooting: the item is not assayed.Repeat assay.The probable cause of the issue is attributed to faulty pressure sensor and solenoid valve.
|
A customer reported error 4017 spec.Sy clog detected on the aia-360 analyzer.The customer performed an all-set-home command; however, the error persisted.An fse was dispatched.A field service engineer (fse) went onsite to address the reported issue, which resulted in the delay of reporting intact parathyroid hormone (ipth) patient results.There was no report of patient intervention or adverse health consequences due to delay in reporting.
|