Model Number AIA-360
Device Problem Obstruction of Flow (2423)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/27/2020
Event Type  malfunction  
Manufacturer Narrative
A field service engineer (fse) was at the customer site to address the reported issue.The fse confirmed and replicated the issue by running qc on the instrument.The fse replaced the sample nozzle; however, the error persisted.The pressure sensor was then replaced and the impasse test result was acceptable but on the high end.Next, the solenoid valve on the sample syringe was replaced which brought the impasse test result near the middle, and the issue was resolved.Instrument validation was performed via quality control (qc); results were within the published ranges.The aia-360 instrument returned to normal operation.No further field action required.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from (b)(6) 2019 to aware date (b)(6) 2020.No other similar complaints were identified during the search period.The aia-360 operator's manual states the following: [4017] spec.Sy clog detected.Description: specimen clog was detected.Troubleshooting: the item is not assayed.Repeat assay.The probable cause of the issue is attributed to faulty pressure sensor and solenoid valve.
 
Event Description
A customer reported error 4017 spec.Sy clog detected on the aia-360 analyzer.The customer performed an all-set-home command; however, the error persisted.An fse was dispatched.A field service engineer (fse) went onsite to address the reported issue, which resulted in the delay of reporting intact parathyroid hormone (ipth) patient results.There was no report of patient intervention or adverse health consequences due to delay in reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-360
Type of Device
AIA-360
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, 10586 23
JA  1058623
MDR Report Key10443100
MDR Text Key206112452
Report Number8031673-2020-00226
Device Sequence Number1
Product Code KHO
Combination Product (y/n)N
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 08/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-360
Device Catalogue Number019945
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-