This is a supplemental report to provide additional information.The subject device was returned to olympus medical systems corp.(omsc) for evaluation.The operation wire was cut at 70 mm from the distal end.The operation wire was broken off at 940 mm from the distal end.The broken surface showed that mechanical tensile load was applied.There was no abnormality in the outer diameter of the three operation wires.The operation wire was broken off at 2000 mm from the distal end.The broken surface showed that mechanical tensile load was applied.There was no abnormality in the outer diameter of the three operation wires.The basket wire was deformed.The manufacturing record was reviewed and found no irregularities.Based on the past similar cases, omsc presumes that the event occurred due to the following occurrence mechanism.1) due to factors such as the quantity, size, shape, hardness of calculus, or strength of the force to close the basket wire, more force than expected was applied to the device when the stone was gripped.2) due to 1), the operation wire near the operation part was broken.3) when the user tried to crush stones using bml-110a-1, due to the factors such as the quantity, size, shape, hardness of calculus, or strength of the force to close the basket wire, more force than expected was applied to the device.4) due to 3), the middle part of the operation wire was broken.The above device handling has warned in the instruction manual.
|