• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RFB VULCAN GENERATOR CE MARK; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RFB VULCAN GENERATOR CE MARK; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 7210812F
Device Problems Noise, Audible (3273); Intermittent Loss of Power (4016)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/04/2020
Event Type  malfunction  
Event Description
It was reported that during a hip arthroscopy, the vulcan generator was making a loud beeping noise and sometimes it would shut off.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3, h6 h10: the reported device, used in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of alarm and power loss could not be reproduced.Unit passed functional testing during 2 hour burn-in and power output was normal throughout testing.All functions perform as expected.All power and impedance readings were within spec.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RFB VULCAN GENERATOR CE MARK
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10446404
MDR Text Key204200402
Report Number1643264-2020-00949
Device Sequence Number1
Product Code GEI
UDI-Device Identifier03596010613561
UDI-Public03596010613561
Combination Product (y/n)N
PMA/PMN Number
K991140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7210812F
Device Catalogue Number7210812F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2020
Initial Date Manufacturer Received 08/04/2020
Initial Date FDA Received08/24/2020
Supplement Dates Manufacturer Received09/22/2020
Supplement Dates FDA Received09/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-