SMITH & NEPHEW, INC. RFB VULCAN GENERATOR CE MARK; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
Model Number 7210812F |
Device Problems
Noise, Audible (3273); Intermittent Loss of Power (4016)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/04/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that during a hip arthroscopy, the vulcan generator was making a loud beeping noise and sometimes it would shut off.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
H3, h6 h10: the reported device, used in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of alarm and power loss could not be reproduced.Unit passed functional testing during 2 hour burn-in and power output was normal throughout testing.All functions perform as expected.All power and impedance readings were within spec.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
|
|
Search Alerts/Recalls
|
|
|