• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVE DIVISION MOSAIC MITRAL BIOPROSTHETIC HEART VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVE DIVISION MOSAIC MITRAL BIOPROSTHETIC HEART VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number 310C27
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Complete Heart Block (2627)
Event Date 07/15/2020
Event Type  Injury  
Manufacturer Narrative
Product analysis: the device remains implanted, therefore no product analysis can be performed.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.Conduction disturbances are known potential adverse effects associated with any cardiac or thoracic procedure and can be resolved with medical treatment or the implant of a permanent pacemaker (with the risk-benefit ratio in favor of implant of the heart valve).A conduction disturbance does not indicate a device malfunction or potential manufacturing issue.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that 4 days post implant of this 27 mm bioprosthetic mitral valve, a permanent pacemaker was implanted.The reason for permanent pacemaker implant was reported as 3rd degree complete heart block.No additional adverse patient effects were reported. .
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIC MITRAL BIOPROSTHETIC HEART VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
MEDTRONIC HEART VALVE DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key10448553
MDR Text Key204206892
Report Number2025587-2020-02670
Device Sequence Number1
Product Code LWR
UDI-Device Identifier00643169594845
UDI-Public00643169594845
Combination Product (y/n)N
PMA/PMN Number
P990064
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/25/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/14/2021
Device Model Number310C27
Device Catalogue Number310C27
Was Device Available for Evaluation? No
Date Manufacturer Received08/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age62 YR
-
-