• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM; EXTERNAL COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM; EXTERNAL COMPONENT Back to Search Results
Model Number DACAPO POWERPACK
Device Problem Fluid/Blood Leak (1250)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device has been explanted and has been returned to innsbruck where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
 
Event Description
A leaking/ broken dacapo battery was received at service and repair.On the repair request form there were battery issues, the battery would not charge.
 
Manufacturer Narrative
Conclusion: the returned device was visually inspected upon arrival.A mechanically damaged housing was observed, most likely due to external mechanical stress.The observed damage did not allow electrical testing to be conducted.The damage to the battery housing was clearly the reason for the malfunction of the device returned by the end-user.This is a final report.
 
Event Description
A leaking/ broken dacapo battery was received at service and repair.On the repair request form there were battery issues, the battery would not charge.The user did not come into contact with the leaking electrolyte.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Type of Device
EXTERNAL COMPONENT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
MDR Report Key10452884
MDR Text Key204344184
Report Number9710014-2020-00470
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P000025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 09/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/26/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDACAPO POWERPACK
Device Lot NumberSN: 1805K149
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2020
Was the Report Sent to FDA? No
Date Manufacturer Received08/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-