• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. MCK FEMORAL-LM-RL-SZ 1; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. MCK FEMORAL-LM-RL-SZ 1; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM Back to Search Results
Model Number 180501
Device Problem Loss of Osseointegration (2408)
Patient Problems Injury (2348); Inadequate Osseointegration (2646)
Event Date 08/15/2019
Event Type  Injury  
Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.Device not returned.
 
Event Description
This pi is for the left knee.As reported in pi (b)(4): it was reported by surgeon that pre op x-rays showed medial compartment wear.All the standard criteria for medial compartment arthroplasty were fulfilled and surgeon performed a bilateral partial knee medial makoplasty.Primary surgery performed on (b)(6) 2017, by the doctor.The surgeon reported that the cementing technique is as what he would do for all medial mako knees and have had no issues.Patient post op x-rays were good position.Even her post op mri done on her knees 9 months after post surgery showed no loosening and stable implants, nor edema.Post op x-rays showed the implants to be perfectly placed.However in 2019 it was noted that she developed accelerated loosening of both medial implants and this resulted in a bilateral revision total knee arthroplasty.The revision was performed by dr on (b)(6) 2019.The operative notes showed bilateral extensive polyethylene wear with resultant extensive synovitis and loosening.Revision performed using non stryker product.No details provided as patient went to consult another surgeon.Updated: 03 aug 2020 - refer implant picture in intake tab.Recently, the patient came back to visit and informed dr.About the explant implant surgery done in year 2019.The surgeon informed stryker sales rep and would like stryker to investigate the possible causing of the loosening.Update: 6 aug 2020: it was informed by the surgeon that the patient want to file a lawsuit against him.Additional information provided by the surgeon: i find this very peculiar.Photos of the retrieved implants show wear of the plastic that is way too advanced for a 2 year post op surgery.I wouldn¿t expect to see this much wear of the poly for that duration.Especially if the implants were well placed on post op x-rays.
 
Event Description
This pi is for the left knee.As reported in pi 2452066 : it was reported by surgeon that pre op x-rays showed medial compartment wear.All the standard criteria for medial compartment arthroplasty were fulfilled and surgeon performed a bilateral partial knee medial makoplasty.Primary surgery performed on (b)(6) 2017, by the doctor.The surgeon reported that the cementing technique is as what he would do for all medial mako knees and have had no issues.Patient post op x-rays were good position.Even her post op mri done on her knees 9 months after post surgery showed no loosening and stable implants, nor edema.Post op x-rays showed the implants to be perfectly placed.However in 2019 it was noted that she developed accelerated loosening of both medial implants and this resulted in a bilateral revision total knee arthroplasty.The revision was performed by dr on (b)(6) 2019.The operative notes showed bilateral extensive polyethylene wear with resultant extensive synovitis and loosening.Revision performed using non stryker product.No details provided as patient went to consult another surgeon.Updated: 03 aug 2020 - refer implant picture in intake tab.Recently, the patient came back to visit and informed dr.About the explant implant surgery done in year 2019.The surgeon informed stryker sales rep and would like stryker to investigate the possible causing of the loosening.Update : 6 aug 2020: it was informed by the surgeon that the patient want to file a lawsuite against him.Additional information provided by the surgeon: i find this very peculiar.Photos of the retrieved implants show wear of the plastic that is way too advanced for a 2 year post op surgery.I wouldn¿t expect to see this much wear of the poly for that duration.Especially if the implants were well placed on post op x-rays.
 
Manufacturer Narrative
Reported event:  an event regarding loosening involving mako femoral component was reported.The event was not confirmed.  method & results:  -device evaluation and results: not performed as product was not returned -clinician review:   a review of the provided medical records and/or x-rays by a clinical consultant indicated: "pi which is from singapore and relates to the right and left knees of the patient respectively.This is a female patient, dob (b)(6) 1960 described as 146 cm tall and weighing 63 kg.Her date of implantation is listed as (b)(6) 2017 and explantation (b)(6) 2019.The event descriptions are identical with pi - 2457135 for the right and pi - 2452066 for the left knee.The event descriptions state: ".Bilateral medial uka.(b)(6) 2017.2019 loosening both medial implants.Bilateral revision tka.Bilateral extensive poly wear.Patient wants to file a lawsuit." (b)(6) 2017 cytopathology report specimen received (b)(6) 2017: "fluid from left knee" "loose bodies of fibrin with chondrogenesis and cartilage." (b)(6) 2016 x-ray ap both knees "erect": mild osteoarthritis both medial compartments with lateral subluxation of both tibia.No clinical or pmh, no operative reports, no examination of explanted components, no post-operative or dated serial x-rays.Based upon the information available for review, neither confirmation of the event descriptions nor preparation of a medical report is possible for this case." -device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.  -complaint history review: there have been no other similar events for the reported lot.  conclusion:  the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.Device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCK FEMORAL-LM-RL-SZ 1
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key10454010
MDR Text Key204387540
Report Number3005985723-2020-00279
Device Sequence Number1
Product Code NPJ
UDI-Device Identifier00848486001438
UDI-Public00848486001438
Combination Product (y/n)N
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 10/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/31/2020
Device Model Number180501
Device Catalogue Number180501
Device Lot Number26490315-01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/30/2020
Initial Date FDA Received08/26/2020
Supplement Dates Manufacturer Received10/02/2020
Supplement Dates FDA Received10/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Weight63
-
-