Model Number LXMC13 |
Device Problem
Patient Device Interaction Problem (4001)
|
Patient Problems
Dysphagia/ Odynophagia (1815); No Code Available (3191)
|
Event Date 07/30/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Date sent: 08/26/2020.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot was not provided; therefore, the manufacturing record evaluation could not be performed.Additional information was requested, and the following was obtained: on what date did the implant take place? (b)(6) 2017.Does the patient have any of the allergies to metals? no.Is the patient currently taking currently taking steroids / immunization drugs? no.Did the patient have any pre-existing dysphagia or other conditions (other than gerd)? no.Was there any hiatal or crural repair done at the same time as the implant? yes.Was mesh used at time of implant? no.What was the reason for removal of the linx device? chronic dysphagia.At the time of removal, was the device found in the correct position/geometry at the time of removal? yes.Have the symptoms resolved since the device was explanted? patient had a lap revision toupet fundoplication at the time of explant.
|
|
Event Description
|
It was reported that during a linx explant, the link was removed.Case was completed.
|
|
Manufacturer Narrative
|
(b)(4).Date sent: 9/29/2020.Device analysis: overall review of the device function and dimensions show no anomalies from a device that has been reasonably changed as part of the explant procedure.Visual analysis was consistent with an explanted device, and link length and tensile force were found to meet the applicable specifications.Overall, no analysis conclusions relevant to the patient experience were found.The lot/batch was not provided; therefore, the manufacturing records could not be reviewed.
|
|
Search Alerts/Recalls
|
|
|