MAUDE Adverse Event Report: TORAX MEDICAL, INC. 1.5T LINX, 13B; ANTI-REFLUX IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORAX MEDICAL, INC. 1.5T LINX, 13B; ANTI-REFLUX IMPLANT Back to Search Results
Model Number LXMC13
Device Problem Patient Device Interaction Problem (4001)
Patient Problems Dysphagia/ Odynophagia (1815); No Code Available (3191)
Event Date 07/30/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).Date sent: 08/26/2020.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot was not provided; therefore, the manufacturing record evaluation could not be performed.Additional information was requested, and the following was obtained: on what date did the implant take place? (b)(6) 2017.Does the patient have any of the allergies to metals? no.Is the patient currently taking currently taking steroids / immunization drugs? no.Did the patient have any pre-existing dysphagia or other conditions (other than gerd)? no.Was there any hiatal or crural repair done at the same time as the implant? yes.Was mesh used at time of implant? no.What was the reason for removal of the linx device? chronic dysphagia.At the time of removal, was the device found in the correct position/geometry at the time of removal? yes.Have the symptoms resolved since the device was explanted? patient had a lap revision toupet fundoplication at the time of explant.
 
Event Description
It was reported that during a linx explant, the link was removed.Case was completed.
 
Manufacturer Narrative
(b)(4).Date sent: 9/29/2020.Device analysis: overall review of the device function and dimensions show no anomalies from a device that has been reasonably changed as part of the explant procedure.Visual analysis was consistent with an explanted device, and link length and tensile force were found to meet the applicable specifications.Overall, no analysis conclusions relevant to the patient experience were found.The lot/batch was not provided; therefore, the manufacturing records could not be reviewed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.5T LINX, 13B
Type of Device
ANTI-REFLUX IMPLANT
Manufacturer (Section D)
TORAX MEDICAL, INC.
4188 lexington avenue north
shoreview MN
MDR Report Key10455124
MDR Text Key205165945
Report Number3008766073-2020-00134
Device Sequence Number1
Product Code LEI
UDI-Device Identifier00855106005332
UDI-Public00855106005332
Combination Product (y/n)N
PMA/PMN Number
P100049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 08/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberLXMC13
Device Catalogue NumberLXMC13
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2020
Initial Date Manufacturer Received 08/03/2020
Initial Date FDA Received08/26/2020
Supplement Dates Manufacturer Received09/17/2020
Supplement Dates FDA Received09/29/2020
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-