Model Number 1192
Device Problem Break (1069)
Patient Problem Inadequate Pain Relief (2388)
Event Date 03/06/2020
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Related manufacturer report: 3006705815-2020-30992, related manufacturer report: 3006705815-2020-30993, related manufacturer report: 1627487-2020-30685.It was reported that patient experienced ineffective stimulation, due to lead migration issue.Upon x-ray observation lead migration was confirmed wherein one lead migrated one vertebral body and the second lead migrated half a vertebral body.Patient denies any traumas or falls.Blood work was done, and no presence of abscess was confirmed.Both leads were explanted, and new leads were implanted.During the procedure, anchor damage issue was observed.Both anchors were explanted, and new anchors were implanted.Effective therapy was established post procedure.Patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWIFT-LOCK ANCHOR
Type of Device
SCS ANCHOR
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key10457839
MDR Text Key204510823
Report Number1627487-2020-30684
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05415067024084
UDI-Public05415067024084
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/26/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/27/2020
Device Model Number1192
Device Catalogue Number1192
Device Lot Number6419262
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/11/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/28/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SCS ANCHOR; SCS LEAD X2
Patient Outcome(s) Other;
Patient Weight50
-
-