Model Number 25 LPR
Device Problems Overheating of Device (1437); Physical Resistance/Sticking (4012)
Patient Problems Burn(s) (1757); Burn, Thermal (2530)
Event Date 07/27/2020
Event Type  malfunction  
Manufacturer Narrative
A short visual inspection and a test run prior to the repair showed that the heat up was reproducible.Root cause has been that the ball bearings have been worn out which caused grinding and as a result vibrations and higher friction and hence heat up.To avoid such issues the user instruction contains already several notes, warnings and requests how to prepare the handpiece for each treatment and how to use it: warning: hazards for the care provider and the patient.In the case of damage, irregular running noise, excessive vibration, untypical warming or when the cutter or grinder cannot be held.Do not use further and notify service.Caution: burning hazard from hot instrument head or hot instruments cover.If the instrument overheats, burns may arise in the oral area.Never contact soft tissue with the instrument head or instrument cover.The following guidelines must be observed to ensure save use of the electrically driven contra-angle handpieces: the service instructions for contra-angle handpieces must be precisely following when using kavo spray or quattrocare care systems.Before each use, the contra-angle handpiece must be checked for external damage.Before each use, perform a test run with the contra-angle handpiece, and watch for atypical heating and unusual noise and vibration.Immediately stop using contra-angle handpieces that act unusual.Never press the push-button during operation.This also includes lifting the cheek or tongue! to ensure proper function, the medical device must be set up according to the reprocessing methods described in the kavo instructions for use, and the care products and care systems described therein must be used.Kavo recommends specifying a service interval at the dental office for a licensed shop to clean, service and check the functioning of the medical device.This service interval depends on the frequency of use and should be adjusted accordingly.(b)(4).
 
Event Description
When product was sent in it was just stated that instrument gets hot and burns patient.Up to now a contact to the dental office was not possible to get further data.
 
Event Description
Finally, it was possible to get a feedback from the dental office.During a dental treatment the head of the handpiece heated up and burned the patient on inner cheek.There was no medical care necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GENTLEPOWER LUX CONTRA ANGLE 25LPR
Type of Device
DENTAL HANDPIECE
Manufacturer (Section D)
KAVO DENTAL GMBH
bismarckring 39
biberach / riss, bw 88400
GM  88400
MDR Report Key10458789
MDR Text Key204526721
Report Number3003637274-2020-00016
Device Sequence Number1
Product Code EGS
Combination Product (y/n)N
PMA/PMN Number
K073478
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 02/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25 LPR
Device Catalogue Number1.007.3580
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2020
Date Manufacturer Received02/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-