It was reported during the initial implant of a 27mm inspiris valve, the physician had difficulty seating the valve.There was a concern the sizer was not an accurate representation for the 27mm valve.The physician sized the patient with an edwards sizing kit.The physician used the replica end of a 27mm sizer which fit the patient.When the 27mm valve was placed in the annulus, the valve would not fit, it was too big.The physician had to remove the valve from the holder and handle to manually place the device.The physician stated the valve seemed about 2mm off from the replica sizer.The physician had to maneuver the valve into place.Echo showed good seating of the valve with no leaks post procedure.Additionally, the physician placed a valsalva tubular graft to address the patients' aortic aneurysm.The patient was not impacted by the manual placement of the valve.
|
H10.Additional manufacturer narrative: added h6 codes.The physician stated the inspiris valve seemed about 2mm off from the replica magna ease sizer.No product was returned for evaluation.As no product, images, or medical records were returned, the complaint is unable to be confirmed and a definitive root cause cannot be conclusively determined.
|