• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. STRAIGHT SCREWDRIVER; SCREWDRIVER, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. STRAIGHT SCREWDRIVER; SCREWDRIVER, SURGICAL Back to Search Results
Model Number 71362293
Device Problem Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/04/2020
Event Type  malfunction  
Event Description
It was reported that the end of driver was found rounded during set up or inspection.No delay or injury reported.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation a visual inspection confirms the straight screwdriver is stripped and rounded off on both ends of the device.The device shows extensive signs of wear/usage.The device was manufactured in 2009.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRAIGHT SCREWDRIVER
Type of Device
SCREWDRIVER, SURGICAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10469243
MDR Text Key205915355
Report Number1020279-2020-04286
Device Sequence Number1
Product Code LRZ
UDI-Device Identifier03596010454188
UDI-Public03596010454188
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 09/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/28/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71362293
Device Catalogue Number71362293
Device Lot Number09DM06909
Date Manufacturer Received08/31/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-