Model Number 7211010 |
Device Problem
Image Display Error/Artifact (1304)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 08/06/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that, the "dyonics 25 control unit" had issues while inserting a new cassette.The unit displayed error message "cassette insertion/pressure error".After this, when tried to solve the issue, it seemed the sensor of insertion was damaged.No case reported; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The reported device, intended for use in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was a relationship found between the returned device and the reported incident.An analysis of the customer provided images found that a cassette insertion error was displayed onscreen and a exposed pressure port from the device.The complaint was confirmed and the root cause was associated with mechanical component failure.Factors, unrelated to the manufacturing and design of the device which could have contributed to the reported event, include pressure transducer malfunction or the latch keeper wear.A review of the device history records could not be completed since the provided product identification information did not match any known release of this part.A complaint history review found concluded this was a repeat issue.
|
|