The customer reported a higher than expected vitros bhcg result on a single non-vitros biorad l1 compared to the vitros peer mean tested on a vitros 5600 integrated system.The assignable cause of the event is unknown.Improper sample handling cannot be ruled out as contributing to the event.It is possible the customer left the uncapped l1 control fluid on board the vitros instrument for >1 hour, although this cannot be confirmed.A vitros bhcg reagent related issue cannot be ruled out as there were insufficient data points to adequately assess vitros bhcg reagent lot 2680.In addition, a vitros instrument related issue did not likely contribute to the event as a vitros bhcg within-run precision test indicated vitros bhcg reagent lot 2680 was performing as expected on the vitros 5600 system.
|