• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIO-DETEK INCORPORATED PRO-PADZ RADIOLUCENT SOLID GEL ELECTRODE, SINGLE; DEFIB/PACING ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIO-DETEK INCORPORATED PRO-PADZ RADIOLUCENT SOLID GEL ELECTRODE, SINGLE; DEFIB/PACING ELECTRODE Back to Search Results
Model Number 8900-4006
Device Problem Arcing of Electrodes (2289)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/01/2020
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that while attempting to cardiovert a (b)(6) male patient, an arc was heard from the electrode pads.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
The pro padz solid gel radiolucent multifunction electrodes were returned to zoll medical corporation opened and adhered together.Visual inspection found a heavy presence of hair on the adhesive of the electrodes and discoloration on the conductive plates indicating the occurrence of discharge.The electrodes appear to be functional although sent to zoll in an unsealed package which dried out the gel.All indications suggest patient preparation was a factor.The electrode pads were scrapped.Zoll recommends that patients are cleaned and hair clipped prior to applying electrode pads to assure good coupling of electrode to skin contact.Poor adherence and/or air under the electrodes can lead to the possibility of arcing and skin burns.This investigation was closed as improper patient preparation prior to the application of the electrodes.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRO-PADZ RADIOLUCENT SOLID GEL ELECTRODE, SINGLE
Type of Device
DEFIB/PACING ELECTRODE
Manufacturer (Section D)
BIO-DETEK INCORPORATED
525 narragansett park drive
pawtucket RI 02861
MDR Report Key10472187
MDR Text Key204985482
Report Number1218058-2020-00066
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00847946016661
UDI-Public00847946016661
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Type of Report Initial,Followup
Report Date 08/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date01/22/2021
Device Model Number8900-4006
Device Catalogue Number8900-4006
Device Lot Number0420A
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Hospital
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/31/2020
Supplement Dates Manufacturer Received08/10/2020
Supplement Dates FDA Received11/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age67 YR
-
-