Related manufacturer report number: 2017865-2020-12580.It was reported that the patient experienced syncope and was dependent on the device for normal function.It was noted during a follow up in clinic that noise, oversensing reproducible with pocket manipulation and arm movements and pause episodes (pacing stopped) were observed on the right ventricular lead and device.A device header anomaly and myopotentials were also suspected and electromagnetic interference (emi) was ruled out.Programming changes were made.The right ventricular lead and device were explanted and replaced.The patient was stable before, during and after the procedure.
|
Final analysis noted the complaint of oversensing, pacing anomaly, and defective device were not confirmed.The device was received from the field in normal working conditions with battery voltage near beginning of life level.Atrial and ventricular sensitivity test under nominal conditions indicated results were normal and within expected sensitivity range.Further evaluation of the output parameter under nominal conditions found all parameters within normal range of operation.Electrical and mechanical tests performed were normal.Additionally, visual inspection of the header and connectors detected no contaminants or foreign material that could contribute to the reported complaints.Longevity assessment was performed, and device was in the normal range of operation with appropriate remaining longevity.
|