SMITH & NEPHEW, INC. JOURNEY ART ISRT ASSM TOOL; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 74018911
Device Problem Material Separation (1562)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/03/2020
Event Type  malfunction  
Event Description
It was reported that, during inspection or set up, the piece in the handle that allows the handle to squeeze has broken off and needs to be replaced.Nothing fell into patient.The procedure was finished using a smith and nephew back up device.No delay to procedure, no harm to patient.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection confirms one of the spring leaf is broken off the device.The broken piece was not returned with the device.The device was manufactured in 2011.The device exhibits signs of significant wear/ usage.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY ART ISRT ASSM TOOL
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10477145
MDR Text Key205121464
Report Number1020279-2020-04331
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010557292
UDI-Public03596010557292
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/01/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number74018911
Device Catalogue Number74018911
Device Lot Number11JM02562
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/12/2020
Date Manufacturer Received08/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-