• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 397002-001
Device Problem Audible Prompt/Feedback Problem (4020)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/05/2020
Event Type  malfunction  
Manufacturer Narrative
The companion 2 driver will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4) initial.
 
Event Description
The companion 2 driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the companion 2 driver was in a store room and not plugged into wall power.When the driver was turned on for a training session, the driver exhibited an emergency battery error alarm.
 
Manufacturer Narrative
The driver's patient file was reviewed and revealed ten emergency battery error alarms thus confirming the customer-reported issue.The root cause was determined to be malfunction of the emergency battery as it was unable to power the driver for the full 10-minute requirement.The inability to power the system for the required duration is a result of loss of battery capacitance.This issue will be monitored and trended as part of the customer experience process.Syncardia has completed its investigation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key10477304
MDR Text Key205493789
Report Number3003761017-2020-00172
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 07/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2021
Initial Date Manufacturer Received 08/06/2020
Initial Date FDA Received09/01/2020
Supplement Dates Manufacturer Received08/06/2020
Supplement Dates FDA Received07/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-