Product event summary: visual inspection of balloon catheter (2af284 with lot # 10013) showed that the catheter was intact with no apparent issues.The contamination issue could not be determined.Smart chip verification indicated the catheter was used for one injection.The balloon catheter passed the performance test and electrical integrity as per specification; impedance was also within specification.The cause of the adverse event could not be determined for this event.In conclusion, the clinical issue (air embolism) occurred during the case.There is no indication of relation of adverse event to the performance of the complaint products.The reported air ingress issue was not confirmed through testing.The balloon catheter passed the returned product inspection as specification.If information is provided in the future, a supplemental report will be issued.
|
It was reported that during a cryo ablation procedure, the patient experienced an air embolism, indicating air ingress occurred.Int ervention was needed in order to exhaust the air from the body.It was also reported the balloon catheter became contaminated at some point in the procedure and needed to be replaced.The case was completed with cryo.No further patient complications have been reported as a result of this event.
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|