Model Number 4FC12
Device Problem Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/23/2020
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the 4fc12 sheath with lot 20166 was returned and analyzed.Visual inspection of the sheath showed the shaft was kinked at 1 and 2.5 inches from the tip, respectively.X-ray inspection didn't show any broken braided wires at the kink locations as well as along the shaft.In conclusion, the sheath failed the returned product inspection due to a sheath kink.If information is provided in the future, a supplemental report will be issued.
 
Event Description
After a completed case, the sheath subsequently tested out of specification per the manufacturer's investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10483265
MDR Text Key209812071
Report Number3002648230-2020-00461
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/11/2021
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number20166
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/31/2020
Initial Date Manufacturer Received 08/18/2020
Initial Date FDA Received09/02/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/12/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age67 YR
-
-