Device evaluated by mfr: the returned product consisted of an ffr comet pressure wire connected to the optical cable connector (occ).The tip, device shaft, sensor port and the coefficient values were examined for damage or any irregularities.The shaft showed 2 kinks located 38cm and 154cm from the tip.There was peeled coating at the 154cm location.The occ handle was connected to the ffr link to verify the signal strength.The signal was present and showed green lights as designed.The wire was inserted into the pressure chamber test equipment and the pressure was increased to verify the sensor was indeed reacting to the pressure increases and decreases.The pressure sensor functioned as designed.The coefficient was confirmed to be programmed per specification.The occ handle was again connected to the ffr link.The device was then connected to the polaris (ilab) test equipment via bluetooth signal.The ffr link and wire communicated to the polaris system and zeroed as designed.The wire was inserted into the test pressure chamber, the wire transferred a pd pressure waveform to the polaris which indicates a functioning wire.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The occ handle cap was loosened to remove the wire.There was no issue with removing the wire.The sensor housing showed no residue of body fluids.Device analysis determined the condition of the returned device was consistent with the reported information.
|