ERIKA DE REYNOSA, S.A. DE C.V. COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
|
Back to Search Results |
|
Model Number 03-2622-3 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/01/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
|
|
Event Description
|
A medwatch form was received via postal mail from an acute user facility at a hospital, which indicated there was a combi set blood leak that occurred during a patient's hemodialysis (hd) treatment.As reported, upon initiation of treatment, blood started leaking from the heparin pigtail line.When the leak was identified, the piece of tubing was ¿lifted at the pigtail area on the line¿ and the entire tubing disconnected from the main line of the dialysis tubing.Blood started coming out at the connector.No patient injury or harm was reported.The reported month of occurrence was provided; however, a specific date was not.Attempts were made to contact both the initial reporter of the event, and the listed contact person at the facility.To date, no additional information has been received, and the sample availability status was not provided.
|
|
Search Alerts/Recalls
|
|
|