• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERIS MEXICO, S. DE R.L. DE C.V. V-PRO 1 PLUS STERILIZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS MEXICO, S. DE R.L. DE C.V. V-PRO 1 PLUS STERILIZER Back to Search Results
Device Problems Fire (1245); Device Emits Odor (1425)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/04/2020
Event Type  malfunction  
Manufacturer Narrative
A technician from steris dealer, device technologies australia (dta), arrived onsite to inspect the unit and found that fittings on the reservoir required replacement.The damage to the fittings allowed sterilant to leak from the reservoir onto electronic components below causing the electrical wires to short and the reported event to occur.The technician performed the necessary repairs, tested the unit, confirmed it to be operating to specification, and returned it to service.The unit was manufactured in 2011 making it approximately 9 years old.A 3-year complaint review indicates this to be an isolated event.No additional issues have been reported.
 
Event Description
The user facility reported that their v-pro 1 plus sterilizer had successfully completed a cycle when user facility personnel noticed a burning smell and observed flames coming from the unit.The fire was extinguished by a user facility employee.No report of injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-PRO 1 PLUS STERILIZER
Type of Device
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, oh 
3927453
MDR Report Key10487827
MDR Text Key206729881
Report Number3005899764-2020-00053
Device Sequence Number1
Product Code MLR
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/07/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-