The field service engineer (fse) conducted a site visit and was able to confirm the reported issue via the error log of 18 flags out of approximately 50 samples.The fse was also able to reproduce the reported issue by running several samples.The fse also checked the main arm sample nozzle and it was clear of any buildup.The fse also inspected the sample nozzle pressure, offset 250, gain s34; within normal range.The fse checked the beltline alignment; tip is centered in sample tube.The fse lubricated sample syringe and sampling assembly mechanics; exercised sample syringe several times in maintenance program.Customer had controls ready to run; all controls in range, no flags.The aia-2000 analyzer is functioning as expected.No further action required by field service.A 13-month complaint history review and service history review for similar complaints were performed for the serial number (b)(4) from 10jul2019 through aware date 10aug2020.There were seven similar complaints including this one that were identified during the searched period.The aia-2000 operator's manual under appendix 4: error messages states the following: error message: [2086] abnormal pressure detected during specimen suction by main arm.Cause: abnormal suction was detected during specimen suction, possibly due to the following causes: the specimen has high viscosity.A half-solid substance floating in the specimen, such as fibrin, was ingested.Air was ingested.Measuring operation is resumed and the measurement result will be flagged (uc) flag.Solution: verify that the specimen is free of solid substances (such as fibrin) or bubbles on the surface and retry the measurement.If the error persists, but the result is judged normal, the specimen viscosity may be high.In this case, the uc flag attached to the result can be deleted later.If the error still cannot be solved, contact tosoh service center or local representatives.The most probable cause of the reported event is lubrication of the main arm assembly and syringe was needed.
|