• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR MITRACLIP SYSTEM STEERABLE GUIDE CATHETER; VALVE REPAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR MITRACLIP SYSTEM STEERABLE GUIDE CATHETER; VALVE REPAIR Back to Search Results
Model Number SGC0301
Device Problems Product Quality Problem (1506); Material Split, Cut or Torn (4008)
Patient Problem No Patient Involvement (2645)
Event Date 08/13/2020
Event Type  malfunction  
Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report torn material.It was reported that during preparation, the distal end of the steerable guide catheter (sgc) felt bumpy.It was then noticed that the flush basin seemed to contain a foreign object, which was potentially some of the hydrophilic coating on the tip of the sgc.The sgc was not used and replaced.There was no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
All available information was investigated and the reported issues of product quality problem-irregular texture on the distal end and hydrophilic coating tip tear could not be confirmed via returned device analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this complaint.Additionally, a review of the complaint history identified no similar incidents for this lot.Based on the information reviewed, a cause for the reported bumpiness on distal end (irregular texture) and potential hydrophilic coating tip tear could not be determined.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of Device
VALVE REPAIR
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10488968
MDR Text Key214617844
Report Number2024168-2020-07354
Device Sequence Number1
Product Code DRA
UDI-Device Identifier08717648216824
UDI-Public08717648216824
Combination Product (y/n)N
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/23/2021
Device Model NumberSGC0301
Device Catalogue NumberSGC0301
Device Lot Number00323U117
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2020
Initial Date Manufacturer Received 08/13/2020
Initial Date FDA Received09/03/2020
Supplement Dates Manufacturer Received10/01/2020
Supplement Dates FDA Received10/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-