This report has been identified as b.Braun medical inc.Internal report number (b)(4).The device logs were obtained and reviewed.The logs show the transfer set is installed and pumping is fairly normal throughout the day.However, additional analysis indicates that the occlusion readings on both the macro and micro drift towards "less force on sensor" as the day goes on.It is theorized that this is an indication that the transfer set appears to have not been clipped-in fully on the right-hand side or became unclipped during compounding (most likely as valve 25 finishes pumping).The occlusion readings sharply drop off showing no force on the sensor.At that time, valve 5 pumps and an occlusion is received because of the unresponsive occlusion sensor.The occlusion signal becomes unresponsive due to the tubing no longer touching the sensor.As a result to no further occlusion reading responsiveness, the user takes the set off and notices valve 25 is turned.This could have happened if the transfer set popped off the right-hand side while valve 25 was closing.There were no alerts because the left-hand side was still clipped in, which is where the magnet which senses the transfer set resides.The transfer set used in this event was not returned for evaluation.However, a transfer set was returned in a similar case under b.Braun internal report (b)(4).Visual examination of that set noted that valve 25 was in an open position.The sample was leak tested per specification and it was noted that air was coming from the end of line 25.It was determined this issue was not a manufacturing defect.A retained transfer set of the same lot was visually inspected and no defects were noted.The retained transfer set was occlusion tested and vacuum leak tested per specification with passing results.Review of the discrepancy management system database performed for the reported lot number did not reveal any abnormalities or non-conformances of this nature.It is possible the user may have improperly latched the manifold onto the compounder and while compounding, the manifold became detached and halted compounding.If additional pertinent information becomes available a follow-up report will be filed.
|