It was reported that the procedure was cancelled.A rotablator rotational atherectomy system console kit was selected for use in a rotational atherectomy procedure.During the procedure, the rotalink plus was connected and the rotation speed was adjusted.However, there was no response from the equipment to the speed adjustment of the control knob leading to high revolutions without control.The procedure was unsuccessful and was cancelled.No patient complications complications were reported.
|
It was reported that the procedure was cancelled.A rotablator rotational atherectomy system console kit was selected for use in a rotational atherectomy procedure.During the procedure, the rotalink plus was connected and the rotation speed was adjusted.However, there was no response from the equipment to the speed adjustment of the control knob leading to high revolutions without control.The procedure was unsuccessful and was cancelled.No patient complications were reported.It was further reported that the operating speed is set according to the size of the burr; a set speed does not come by default.The maximum speed reached is 260 rate per minute (rpm).The operating speed could not be established as the equipment did not respond to the rpm setting.To complete the procedure, high pressure balloons were used to modify the plate and implant the stent.The patient is stable and on medical discharge.
|