• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK STOPCOCK AND I.V. SOLUTION ADMINISTRATION SETS; STOPCOCK, I.V. SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK STOPCOCK AND I.V. SOLUTION ADMINISTRATION SETS; STOPCOCK, I.V. SET Back to Search Results
Catalog Number 2C8607
Device Problem Connection Problem (2900)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/07/2020
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a clearlink system extension set had a luer adapter that would not "slide up to screw on to what it needed to connect to".This issue was identified during patient use.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H10: the actual sample was received for evaluation.Visual inspection was performed using the naked eye which did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed including clear passage and pressure testing; the device performed according to product specifications.Additional pull test was performed and no separation issues noted.The luer lock however was stuck to its male adapter.The reported condition was verified.The cause of the condition was not determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK STOPCOCK AND I.V. SOLUTION ADMINISTRATION SETS
Type of Device
STOPCOCK, I.V. SET
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10491707
MDR Text Key205899302
Report Number1416980-2020-05453
Device Sequence Number1
Product Code FMG
UDI-Device Identifier00085412071671
UDI-Public(01)00085412071671
Combination Product (y/n)Y
PMA/PMN Number
K130245
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 10/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C8607
Device Lot NumberDR20C04028
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2020
Initial Date Manufacturer Received 08/10/2020
Initial Date FDA Received09/03/2020
Supplement Dates Manufacturer Received09/28/2020
Supplement Dates FDA Received10/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-