MPRI ATTAIN PERFORMA STRAIGHT MRI SURESCAN; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE Back to Search Results
Model Number 439878
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Perforation of Vessels (2135); Cardiac Tamponade (2226); Pericardial Effusion (3271)
Event Date 08/07/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a competitive device upgrade procedure, while implanting the left ventricular (lv) lead, the lead could not be advanced far enough along the posterior lateral cardiac vein so it was decided to attempt via the middle cardiac vein.Contrast was delivered and a perforation of the middle cardiac vein was noted.Additional information reported that the patient experienced a pericardial effusion which resulted in tamponade.A pericardiocentesis was performed and a drain was placed.The lead was removed and the implant attempt was abandoned.The patient will be implanted with an epicardial lead at a later date.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN PERFORMA STRAIGHT MRI SURESCAN
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10492008
MDR Text Key205652662
Report Number2649622-2020-16936
Device Sequence Number1
Product Code OJX
UDI-Device Identifier00643169601840
UDI-Public00643169601840
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P080006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 09/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/03/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/13/2020
Device Model Number439878
Device Catalogue Number439878
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/07/2020
Date Device Manufactured10/12/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age67 YR
-
-