No product has been returned for evaluation nor were radiographs or images provided to confirm the alleged event.The root cause could not be determined, however review of the provided information indicates incomplete trialing/space prep or implant selection may be a possible cause or contributors.Labeling review: ".Potential adverse events and complications: potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s).".".Warnings, cautions and precautions: the subject device is intended for use only as indicated.The implantation of spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.Correct selection of the implant is extremely important.The potential for success is increased by the selection of the proper size of the implant.".
|