Model Number 71430415
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 08/10/2020
Event Type  malfunction  
Event Description
It was reported that the gii ps hi flex isrt tr s5-6 9 was found to be gouged and cracked during a routine efip inspection.There were no reported problems from any cases.
 
Manufacturer Narrative
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device is worn and damaged with chips and gouges, rendering the device inoperable.The device was manufactured in 2010 and shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PS HI FLEX ISRT TR S5-6 9
Type of Device
TRAY, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10497571
MDR Text Key205872856
Report Number1020279-2020-04413
Device Sequence Number1
Product Code FSM
UDI-Device Identifier03596010493057
UDI-Public03596010493057
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 11/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71430415
Device Catalogue Number71430415
Device Lot Number10MM11395
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 08/10/2020
Initial Date FDA Received09/04/2020
Supplement Dates Manufacturer Received11/06/2020
Supplement Dates FDA Received11/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-