Model Number AIRLIFE |
Device Problem
Chemical Problem (2893)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/19/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
At this time, the suspect device has not been returned for evaluation.No root cause has been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information was received.
|
|
Event Description
|
The customer reported that the 124030eu airlife® filtered small volume nebulizer system experienced much longer therapy (nebulizing pentamidine) than they used to be.Filter lets through medication which are irritating for clinician.Also "white smoke" which came out of the filter that irritated their eyes after treatment.The customer confirmed that there was no harm or hazard occurred to the patient.They still continued even though it looked longer and even there was emanations in the room.No intervention was done to the patient.
|
|
Manufacturer Narrative
|
Result of investigation: the vyaire failure analysis laboratory received the two open samples from p/n 124030eu without lot number for the investigation.A visual inspection was performed and no issues were found the two samples.Therefore, reported defect was not confirmed.The root cause is cannot be confirmed.
|
|
Search Alerts/Recalls
|