Model Number CI-1601-05 |
Device Problems
Malposition of Device (2616); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/04/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Event Description
|
The recipient is reportedly experiencing no sound due to an incorrect electrode position.A ct scan confirmed the incorrect electrode position.There was no difficulties with initial electrode insertion reported.Revision surgery will be scheduled.
|
|
Manufacturer Narrative
|
On (b)(6) 2020, the recipient underwent repositioning surgery.The recipient's activation went well and the issues resolved.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|