• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 397002-001
Device Problem Output Problem (3005)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/27/2020
Event Type  malfunction  
Manufacturer Narrative
The companion 2 driver will be evaluated by syncardia.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
While performing a routine evaluation, a syncardia technician reported that the companion 2 driver functional test results were outside of the acceptance criteria.
 
Manufacturer Narrative
The reported issue was reproduced during investigation testing.The root cause of the left driveline pressure being out of specification was determined to be a malfunction of the left electronic pressure regulator.Syncardia has a corrective and preventive action (capa) to address electronic pressure regulators.Syncardia has completed its investigation and is closing this file.Ce 5360 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key10502766
MDR Text Key205937964
Report Number3003761017-2020-00195
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 10/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 08/27/2020
Initial Date FDA Received09/08/2020
Supplement Dates Manufacturer Received08/27/2020
Supplement Dates FDA Received10/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-