MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 595000-001
Device Problems No Apparent Adverse Event (3189); Audible Prompt/Feedback Problem (4020)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/26/2020
Event Type  malfunction  
Manufacturer Narrative
The freedom driver will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.Ce (b)(4) initial.
 
Event Description
The customer, a syncardia certified hospital, reported that the freedom driver exhibited a fault alarm while supporting a patient.There was no reported adverse patient impact.The customer also reported that the patient was subsequently switched to a backup freedom driver.
 
Manufacturer Narrative
The driver's alarm history was reviewed and revealed a "49" alarm code which occurred in the field and was most likely the alarm experienced by the patient.One of the conditions that can produce this alarm is patient hypertension or hypervolemia.The customer did report that there was concern the patient was hypertensive at the time of the driver alarm.The driver passed all sections of functional testing, including a 48-hour observation run with no alarms or abnormalities.In an attempt to replicated the customer conditions, a valsalva maneuver test at both normotensive settings and hypertensive settings was performed on the driver.During these tests, the driver performed as intended annunciating an alarm due to low cardiac output ([<=3.5 lpm) and recovering after a few seconds when cardiac output increased above 3.5 lpm, however, no permanent alarms were recorded.The driver performed as intended with no evidence of a device malfunction.The root cause of the customer-reported fault alarm could not be conclusively determined but was most likely due to the patient conditions, i.E.Hypertensive.This issue will be tracked and trended as part of the customer complaint process.Syncardia has completed its investigation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson, az
Manufacturer Contact
sagar pimpalwar
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key10503227
MDR Text Key206129442
Report Number3003761017-2020-00188
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/26/2020
Initial Date FDA Received09/08/2020
Supplement Dates Manufacturer Received08/26/2020
Supplement Dates FDA Received09/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/30/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age45 YR
Patient SexMale
-
-