• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-ASD-024
Device Problem Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/17/2020
Event Type  malfunction  
Manufacturer Narrative
Further information regarding this event has been requested.Investigation results will be provided in a subsequent submission.
 
Event Description
On (b)(6) 2020 a 24mm asd device (9-asd-024 lot# 6905183) was chosen for procedure.During the procedure it was discovered by the user to deformed, reporting a "cobra head" shape.A new 24mm device was prepared and used to successfully complete the surgery without incident.(9-asd-024 lt#7101837).
 
Manufacturer Narrative
Additional information: d10,g4, h2, h3, h6 & h10.The reported event of device deformation could not be confirmed.The investigation confirmed the device met visual and functional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
MDR Report Key10504745
MDR Text Key213684528
Report Number2135147-2020-00396
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00811806010199
UDI-Public00811806010199
Combination Product (y/n)N
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Remedial Action Replace
Type of Report Initial,Followup
Report Date 09/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2024
Device Model Number9-ASD-024
Device Catalogue Number9-ASD-024
Device Lot Number6905183
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/17/2020
Initial Date FDA Received09/08/2020
Supplement Dates Manufacturer Received09/17/2020
Supplement Dates FDA Received09/18/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age50 YR