SMITH & NEPHEW, INC. JOURNEY DCF AP FEM CUT BLK 5; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 74012415 |
Device Problem
Break (1069)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/14/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that either the spring or the screw is broken on this cutting block.It is unknown when this happened or if there was a case involved.
|
|
Manufacturer Narrative
|
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection confirmed the stated failure mode.The screw within the device is damaged rendering the device inoperable.The device was manufactured in 2017 and shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
|
|
Search Alerts/Recalls
|
|
|