BOSTON SCIENTIFIC CORPORATION PROMUS ELITE; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number 10675
Device Problem Failure to Advance (2524)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/12/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluation by mfr: p elite ous mr 28 x 2.50mm stent delivery system was returned for analysis.A visual examination of the stent found signs of stent damage.Stent struts from the most proximal strut row were lifted and pulled distally.The undamaged stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple kinks along the length of the hypotube shaft.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
 
Event Description
Reportable based on device analysis completed on 17aug2020.It was reported that crossing difficulties were encountered.Vascular access was obtained via the femoral artery.The target lesion containing a <=45 degrees bend was located in the mildly tortuous and mildly calcified left circumflex artery.A 28 x 2.50 promus elite mr drug-eluting stent was advanced but failed to cross the lesion.The procedure was completed with another of the same device.No patient complications were reported and the patient's condition was stable.However, returned device analysis revealed stent damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS ELITE
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key10508955
MDR Text Key207539621
Report Number2134265-2020-11382
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 09/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/29/2021
Device Model Number10675
Device Catalogue Number10675
Device Lot Number0024248037
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/17/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/30/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-