Model Number 2544-01-015 |
Device Problems
Break (1069); Device-Device Incompatibility (2919); Mechanical Jam (2983)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Code Available (3191)
|
Event Date 08/24/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that the cr and ps extension spacer block was stuck to the dog bone.Surgeon could not remove it to put on flexion spacer block.Trying to separate it, the surgeon managed to separate it from the dog bone with an osteotome.In doing so, the spacer block broke in 3 large pieces.None of the pieces landed in the patient.The pieces were thrown in the sharps container by a nurse and were unable to retrieve.No surgical delay.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned for evaluation.The investigation could not draw any conclusions about the reported event without the device to examine.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
|
|
Search Alerts/Recalls
|