Model Number PL572T |
Device Problems
Mechanical Problem (1384); Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/10/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation on going.Additional information / investigation results will be provided in a supplemental report.
|
|
Event Description
|
It was reported that there was an issue with ligature clip.It was reported that the clip magazine could not be attached to the applying forceps.There was no patient harm.Additional information was not provided nor available / was not available.Additional patient information is not available.The adverse event / malfunction is filed under aag reference (b)(4).
|
|
Manufacturer Narrative
|
Investigation results: visual investigation: the investigation was carried out visually and microscopically.During the visual inspection of the product, we found a broken latch.The surface of the breakage does not show any anomalies.It was found that one clip is missing and there are still 11 clips in the magazine.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.Based upon the investigations results a capa is not necessary.
|
|
Event Description
|
The adverse event / malfunction is filed under aag reference: (b)(4).
|
|