• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG LIGATURE CLIP 12 MAG.= 144 PCS.; LAPAROSCOPIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG LIGATURE CLIP 12 MAG.= 144 PCS.; LAPAROSCOPIC SURGERY Back to Search Results
Model Number PL572T
Device Problems Mechanical Problem (1384); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/10/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation on going.Additional information / investigation results will be provided in a supplemental report.
 
Event Description
It was reported that there was an issue with ligature clip.It was reported that the clip magazine could not be attached to the applying forceps.There was no patient harm.Additional information was not provided nor available / was not available.Additional patient information is not available.The adverse event / malfunction is filed under aag reference (b)(4).
 
Manufacturer Narrative
Investigation results: visual investigation: the investigation was carried out visually and microscopically.During the visual inspection of the product, we found a broken latch.The surface of the breakage does not show any anomalies.It was found that one clip is missing and there are still 11 clips in the magazine.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.Based upon the investigations results a capa is not necessary.
 
Event Description
The adverse event / malfunction is filed under aag reference: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGATURE CLIP 12 MAG.= 144 PCS.
Type of Device
LAPAROSCOPIC SURGERY
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key10509950
MDR Text Key207343749
Report Number9610612-2020-00502
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
PMA/PMN Number
K081031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPL572T
Device Catalogue NumberPL572T
Device Lot Number52596627
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2020
Date Manufacturer Received10/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-