(b)(4).Batch #u5aj2j.Investigation summary: the analysis found that one pcee60a device was returned with no apparent damage and with a gst60t reload present.The reload was received fully fired, with the pan detached from the reload.The device was tested for functionality in the straight position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples met the staple release criteria.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformance were identified.It is also possible that handling by the customer could dislodge the pan.Dislodging as a result of reload removal from the device is the most likely scenario.Please reference the instruction for use for more information.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Complaint information will be included in complaint trending that is reviewed on a regular basis to determine if further action is necessary.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|