Model Number PCEE60A
Device Problem Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/24/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch #u5aj2j.Investigation summary: the analysis found that one pcee60a device was returned with no apparent damage and with a gst60t reload present.The reload was received fully fired, with the pan detached from the reload.The device was tested for functionality in the straight position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples met the staple release criteria.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformance were identified.It is also possible that handling by the customer could dislodge the pan.Dislodging as a result of reload removal from the device is the most likely scenario.Please reference the instruction for use for more information.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Complaint information will be included in complaint trending that is reviewed on a regular basis to determine if further action is necessary.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that during a thoracoscopic pneumonectomy, the knife did not move forward during use.The sled of the cartridge moved forward slightly.The cartridge color was gold.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERED 60 ECHELON +, 280MM SHAFT
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 00969
*   00969
6107428552
MDR Report Key10510855
MDR Text Key207229913
Report Number3005075853-2020-04545
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10705036015017
UDI-Public10705036015017
Combination Product (y/n)N
PMA/PMN Number
K110385
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Other
Type of Report Initial
Report Date 06/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/30/2023
Device Model NumberPCEE60A
Device Catalogue NumberPCEE60A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2020
Date Manufacturer Received08/10/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/19/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-