Model Number 71364009
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 08/17/2020
Event Type  malfunction  
Event Description
It was reported that during set up/inspection the femoral head impactor tip was found in two pieces.No case related therefore no patient.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection confirms the black femoral head is fractured into two pieces.Both pieces were returned with the device.This device shows signs of significant wear/usage.This device was manufactured in 2011.A review of complaint history on the listed part revealed no prior complaint for the listed batch with the same failure mode.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL HEAD IMPACTOR
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10511319
MDR Text Key206284368
Report Number1020279-2020-04490
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010199256
UDI-Public03596010199256
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71364009
Device Catalogue Number71364009
Device Lot Number11DM02151
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2020
Initial Date Manufacturer Received 08/17/2020
Initial Date FDA Received09/09/2020
Supplement Dates Manufacturer Received09/12/2020
Supplement Dates FDA Received09/18/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-