Model Number 3822 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/03/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that the balloon leaked.A 10/3.50 flextome cutting balloon was selected for use.During the procedure, it was noted that the balloon was leaking liquid when inflating.The balloon was simply pulled out from the patient's body and the procedure was completed with another of the same device.No complications reported and patient was stable post procedure.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: the device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the device was subjected to positive pressure.The returned device was attached to an encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a balloon pinhole located at the proximal markerband.An examination of the balloon material and proximal markerband identified no issues which could potentially have contributed to this complaint.The rated burst pressure for this device is 12 atmospheres.No issues were noted with the tip section of the device.A visual and microscopic examination found no issue with the markerbands.A visual and tactile examination found no issues or damage to the shaft or hypotube of the device.No other issues were identified during the product analysis.
|
|
Event Description
|
It was reported that the balloon leaked.A 10/3.50 flextome cutting balloon was selected for use.During the procedure, it was noted that the balloon was leaking liquid when inflating.The balloon was simply pulled out from the patient's body and the procedure was completed with another of the same device.No complications reported and patient was stable post procedure.
|
|
Search Alerts/Recalls
|